Fluke 9190A Ultra-Cool Field Metrology Well

Key features
Ultra-cool dry-block calibrator with best-in-class stability

The Fluke 9190A Ultra-Cool Field Metrology Well is the most accurate and stable, cold temperature dry-block on the market. It’s ideal for applications that demand strict quality control and regulatory process compliance. These applications include on-location validation and calibration of RTDs, thermocouples, thermometers, and sensors used with process control equipment such as medical freezers, laboratory refrigerators, cold rooms, blood banks, sterilizers (autoclaves), and freeze dryers.

Wide temperature range
–95 °C to 140 °C

Excellent accuracy
Accuracy using built-in referencethermometer readout: ±0.05 °C full range
Display accuracy: ±0.2 °C full range
Best-in-class stability
±0.015 °C full range

Fast cooling time
23 °C to –90 °C: 80 minutes
23 °C to –95 °C: 90 minutes
140 °C to 23 °C: 60 minutes

Portability
Weighs only 16 kg (35 lbs)
Built-in front and back handlesfor easy two-handed carry
Best measurement practices
Conforms with EURAMET cg-13 guidance on measurement practices for temperature calibrators
Download
Produk Lainnya
Fluke Calibration 6531 6532 EDWT Electronic Deadweight Tester Kits
Fluke Calibration 6531, 6532 E-DWT Electronic Deadweight Tester Kits
Fluke Calibration electronic deadweight tester kits feature E-DWT-H configurations combined with the accessories needed for a complete calibration system. Model 6531 options feature E-DWT-H configurations with a single quartz reference pressure transducer (Q-RPT) to offer percent of reading performance from 10 % to 100 % of device full scale. Model 6532 options feature an additional Q-RPT to maximize workload coverage, with percent of reading performance from 1 % to 100 % of device full scale. An included vacuum fill kit allows the E-DWT to be filled with your choice of compatible fluid, while eliminating problematic air pockets from the test circuit. Also included is a test station adapter to mount just about any pressure device to be tested without using PTFE tape or wrenches. 6531 and 6532 kits provide all you need to perform high level hydraulic pressure calibration right out of the box.
• Pressure ranges available to 200 MPa (30,000 psi)
• ± 0.02 % of reading total one-year accuracy, 0.025 % for two years
• Low torque variable volume allows for pressure generation and control up to 200 MPa (30,000 psi) with minimal physical effort
• Built-in priming system to fill system with test fluid and purge unwanted air to assure smooth operation
• Unique test port design eliminates the need for PTFE tape or wrenches. Adapters included for 1/8 in, 1/4 in, 3/8 in and 1/2 in NPT, 1/8 in, 1/4 in, 3/8 in and 1/2 in BSP, M20 x 1.5 and M14 x 1.5
• Compatible with a wide range of liquids, including water. Vacuum fill kit included to fill the E-DWT with any compatible test medium
• AutoTest™, AutoRange, ready/not ready indicator, onboard data capture and storage, overpressure alarm, RS-232 remote interface and other advanced features to simplify testing, improve safety and prevent calibration errors and damage to equipment
• Optional foot switch accessory allows hands-free data collection when running AutoTests
• Rechargeable battery pack and transport case options for field operation

Key Features
• Eliminates the need for weights and piston-cylinder changes.
• Real-time pressure indication without need for corrections.
• Set and read any pressure value exactly, operates in any unit of measure, and can switch easily between units.
• Offers ± 0.02% of reading total one-year measurement uncertainty.
Fluke 9009 Industrial DualBlock Thermometer Calibrator
Fluke 9009 Industrial Dual-Block Thermometer Calibrator
Double your productivity or cut your calibration time in half
• Temperatures from –15 °C to 350 °C in one unit
• Two wells in each block for simultaneous comparison calibrations
• Rugged, lightweight, water-resistant enclosure

You’ve been asking for it and now we’re making it for you. The Model 9009 Industrial Dual-Block Calibrator lets you calibrate at hot and cold temperatures at the same time. Double your productivity or cut your calibration time in half—either way you look at it, your in-field temperature calibrations just got easier.
Fluke 700HTP2 Hydraulic Test Pump
Fluke 700HTP-2 Hydraulic Test Pump
The Fluke 700HTP-2 is a hand operated hydraulic pressure pump designed to generate calibration pressures up to 10,000 psi/690 bar. Designed for use with Fluke-700 series high pressure modules such as the Fluke-700P09, 700P29, 700P30 and 700P31.
Ideal for calibration when connected to a Fluke high pressure module via the reference port. The 700HTP-2 uses distilled water or mineral-based hydraulic oil as a source for calibration of high pressure input devices such as pressure transmitters or pressure gauges. Can also be used with Fluke 717 pressure calibrators.

Fitur utama
• Variable size stroke adjustment, allows the user to vary the volume and pressure per stroke
• Integral pressure adjustment vernier knob for fine pressure adjustment
• Directly connection to pressure modules consolidates two separate pieces into one when performing pressure calibration and testing
• Two ¼ inch NPT pressure ports, one for reference (pressure module or gauge), one for UUT
• Compatible with optional Fluke-700PRV pressure relief valve designed to protect pressure modules from overpressure damage
• Improved priming
• Improved stroke limiting mechanism
Note: The user must provide a hose with appropriate end fittings to apply pressure from the UUT port to the calibration unit under test such as the Fluke-700HTH-1 hose kit
Fluke 4180 4181 Precision Infrared Calibrators
Fluke 4180, 4181 Precision Infrared Calibrators
Accuracy and traceability for infrared measurements

Now it’s easy to increase your IR temperature measurement accuracy with the new 4180, 4181 Precision Infrared Calibrators from Fluke Calibration. These infrared temperature calibrators give more consistent, accurate and reliable calibrations because emissivity is radiometrically calibrated, and the target size minimizes size of source effect errors. In addition, these infrared calibrators simplify calibration because they uniquely compensate for errors caused by thermometer emissivity settings.

New accredited performance for point-and-shoot calibrations
• Calibrated radiometrically for meaningful, consistent results.
• Accredited calibration included
• Accurate, reliable performance from –15 °C to 500 °C
• Large target size of 152 mm (6 in) is required for calibrating most thermometers.
• Light instrument weight of 8.6 kg (19 lbs) makes it easy to lift and carry.
• Intuitive, easy-to-read display that indicates when temperature is stable.